Welch Allyn PIC 50 Automated External Defibrillators
Published: Fri Jul 07 00:00:00 -0400 2006
MRL, Inc., and FDA notified healthcare professionals of a Class I recall of Welch Allyn PIC 50 Automated External Defibrillators, catalog #97108X manufactured from March 2002 through October 2004. An electrical contact problem may result in the device's failure to provide a defibrillation shock, that could result in delay or failure to resuscitate the patient. This failure may be accompanied by various error messages on the display panel, including the "Defib Comm" error message.
Related MedWatch Alerts:
Welch Allyn AED20 Automatic External Defibrillators