Hamilton Medical, Inc. RAPHAEL Ventilators

Published: Mon Jun 12 00:00:00 -0400 2006

Hamilton Medical, Inc. and FDA notified healthcare professionals of a recall of certain RAPHAEL model ventilators with older generation software, due to a software algorithm designed to suppress false positive alarms that may preclude any alarm (no visible or audible alarms are triggered). The following RAPHAEL ventilators are affected: RAPHAEL (Software version 2.2x) RAPHAEL Silver (Software version 2.2xS) RAPHAEL Color (Software version 2.2xC, 2.2xCU)